4. Remove diaphragm from armature by pulling it
over the retaining screw.
5. Replace diaphragm. –Wet new diaphragm hole
liberally with water.
6. Pop new diaphragm onto armature screw post.
7. Wet with water inside wall and lip of diaphragm.
8. Press diaphragm onto valve housing by pressing
armature down.