AM-5150 ENMET Corporation
3
2.3 Circuit Board Features
The Display Panel is attached by a cable and is released by unscrewing the 4 screws located in the corners. After releasing the
panel, it is swung upward, exposing the interior of the enclosure. The Circuit Board is mounted at the back surface of the
enclosure interior. Features are shown in Figure 2.
Feature Description
Relay Terminals:
J14, J15, J16, J17
This group of terminals is located on the Circuit Board.
For the contacts for each of three alarm relays, and for the contacts of a fault relay.
See Section 3.3
Terminal J12
For V
DC
back-up power in and the 4-20 mA output. See Section 3.1.2
Sensor Terminal J8
Sensor connection, See Section 3.2
Data Terminal J19
RS-485 input/output
Potentiometer 4
Sensor heater voltage adjustment, See Section 3.2.1
Potentiometer 2 & 3
Not used in AM-5150 Do Not Adjust
Figure 2: AM-5150 Circuit Board Features
J14, J15, J16, J17
K1, K2, K3, K4
DC Power In
Terminal J19
Adjustment POT 4
POT 2
Terminal J8