Enmet EX-6165 Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messgeräte Enmet EX-6165 herunter. ENMET EX-6165 User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - ENMET Corporation

ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 www.enmet.com Manual Part Number 80003-123 April 2013

Seite 2 - Table of Contents

EX-6120, 6165 SERIES IR ENMET Corporation 7 3.8 Initial Gas Testing EX-6120, 6165 SERIES IR units are factory calibrated as detailed on the calibrati

Seite 3 - List of Figures and Tables

EX-6120, 6165 SERIES IR ENMET Corporation 8 4.0 Operation The EX-6120, 6165 SERIES IR incorporates magnetically operated switches to allow non-intrus

Seite 4 - 1.0 Introduction

EX-6120, 6165 SERIES IR ENMET Corporation 9 5.0 Calibration 5.1 Sensor Calibration This feature allows the sensor to be calibrated. Ensure that the

Seite 5 - IR Features

EX-6120, 6165 SERIES IR ENMET Corporation 10 6 Sensor Replacement WARNING: Power must be removed from the EX-6120, 6165 SERIES IR before this or any

Seite 6 - IR Mounting

EX-6120, 6165 SERIES IR ENMET Corporation 11 Step 3 Remove the sensor by pulling downwards. Step 4 Carefully locate the pins in the sockets then

Seite 7 - IR to a Control Unit

EX-6120, 6165 SERIES IR ENMET Corporation 12 Step 5 Replace the sensor housing by rotating in a clockwise direction until fully tightened. Step 6

Seite 8 - Connect enclosure

EX-6120, 6165 SERIES IR ENMET Corporation 13 7 Fuses A 1Amp anti-surge fuse, housed within a surface mounted holder, is located within the EX-6120,

Seite 9 - 3.7 Indicators

EX-6120, 6165 SERIES IR ENMET Corporation 14 9 Certification The EX-6120, 6165 SERIES IR(FGD10 Series) of Gas Detectors are ATEX and IECEx certified

Seite 10 - 3.8 Initial Gas Testing

EX-6120, 6165 SERIES IR ENMET Corporation 15 11 Menu Mode The EX-6120, 6165 SERIES IR uses a password system to restrict the end user from carrying o

Seite 11 - 4.0 Operation

EX-6120, 6165 SERIES IR ENMET Corporation 16 11.2.6 E : 6 – Decimal Places The number of decimal places that are used to display the detected gas lev

Seite 12 - 5.0 Calibration

Table of Contents 1.0 INTRODUCTION ...

Seite 13 - 6 Sensor Replacement

EX-6120, 6165 SERIES IR ENMET Corporation 17 11.2.11 E : 13 – Relay 1 The unit is fitted with a relay that is operated in conjunction with the alarm

Seite 14

EX-6120, 6165 SERIES IR ENMET Corporation 18 11.2.16 E : 20 – Negative Zero Suppression This option is used to allow the user to suppress small amoun

Seite 15 - Set Screw

EX-6120, 6165 SERIES IR ENMET Corporation 19 Appendix A A.1 Analog Output Calibration The analogue output is that of a current source. In order to ca

Seite 16 - 8 Specifications

EX-6120, 6165 SERIES IR ENMET Corporation 20 A.3 Communications Protocol The communications protocol used by the EX-6120, 6165 SERIES IR is used for

Seite 17 - 9 Certification

EX-6120, 6165 SERIES IR ENMET Corporation 21 A.4.2 Reading a Variable Send a read frame with the Variable ID to be read: DLE RD var-id DLE EOF C

Seite 18 - 11 Menu Mode

EX-6120, 6165 SERIES IR ENMET Corporation 22 WARRANTY ENMET warrants new instruments to be free from defects in workmanship and material under norm

Seite 19 - IR Sensor Selections

PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instrumen

Seite 20 - 11.2.12 E : 14 – Relay 2

Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: 734.761.1270 FAX Number: 734.

Seite 21

2 List of Figures and Tables FIGURE 1: EX-6120, 6165 SERIES IR FEATURES ...

Seite 22 - Appendix A

ENMET Corporation EX-6120, 6165 SERIES IR 1 1.0 Introduction The ENMET EX-6120, 6165 SERIEs IR Fixed Gas Detectors for the detection of Flammable g

Seite 23 - A.4 Frame Structure

EX-6120, 6165 SERIES IR ENMET Corporation 2 2.0 Features of the EX-6120, 6165 SERIES IR The EX-6120, 6165 SERIES IR is an ATEX and IECEx certified fi

Seite 24 - A.4.4 Writing a Variable

EX-6120, 6165 SERIES IR ENMET Corporation 3 3.0 Installation of the EX-6120, 6165 SERIES IR CAUTION: Area must be declassified during installation. T

Seite 25 - Notes:

EX-6120, 6165 SERIES IR ENMET Corporation 4 Important Connect enclosure earth-ground to a high integrity earth ground ATEX certified Ex d armoured ca

Seite 26

EX-6120, 6165 SERIES IR ENMET Corporation 5 Figure 5: Unplugging the Display Board for Connection Purposes 3.2 Power Supply The EX-6120, 6165 SERIES

Seite 27 - Repair Return Form

EX-6120, 6165 SERIES IR ENMET Corporation 6 3.5 Following Completion of the Installation: a) Replace the top display board using the locating pins o