ENMET Corporation OMNI-4000 S/N 6705 and above
42
Selecting the Maintenance Menu
With the device turned on, plug in the DIN plug. See Figure 3-5. The display reads "Program a sensor" signaling
access to the main maintenance menu.
Figure 3-5: Position of DIN Plug
NOTE: Insufficient battery voltage prevents access to the main menu. The operator is warned with a "Maintenance
impossible" message. Maintenance operations can be continued either by connecting the charger or by changing the
battery pack.
Selecting the Calibration Menu
The display reads:
Press once the MENU switch. The display now reads:
Confirm the choice by pressing the ENTER switch. The display reads; for example:
Program a
sensor
Calibrate a
sensor
Select
sensor: H2S
DIN PlugDIN Plug