3.7 Replacing the Oxygen Sensor
This procedure is used when a new oxygen sensor is installed.
Equipment Required
A small phillips head screwdriver
A new oxygen sensor
Replacing the Oxygen Sensor
Use the screwdriver to remove the cover which contains the sensor openings. See Figure 3-14
Remove the old oxygen sensor
Install the new sensor. The connector ensures a correct fitting. Be careful not to damage the male part of the
connector.
Put the cover protecting the sensors back into place
Figure 3-14: Acceptable Locations of Oxygen and Toxic Sensors
Possible locations of the oxygen sensor are channels 2, 3, and 4. See Figure 3-14.
Checking the Oxygen Sensor
In the interests of safety, check the sensor by:
Pressing the ON/OFF switch
Checking that the displayed reading is 20.9%, in a normally ventilated environment. If it is not, attempt on auto-
zero or calibration. If this doesn't work, replace the sensor.
Explosive Gas
Sensor
CO2 Sensor
(if needed)
Toxic Gas or
Oxygen Sensor
Locations