9
5.3 Ending a Schedule
If a schedule is running the instrument must be turned off first, then turned on again by pressing:
POWER, SAMPLE and CAL ▲ at the same time, the screen will display:
ConF briefly, followed by
Schy which means Schedule YES
Press arrows to select Schn which means Schedule NO
Press the SAMPLE button to select then hold down the POWER button.
Schedule Mode is now off.
5.4 Stop On Full Mode
If Stop On Full Mode is on, and the internal FRAM
htV-m
memory is full, the
htV-m
will not sample until memory is cleared.
If Stop On Full Mode is off, and the internal FRAM
htV-m
memory is full, the oldest readings will be overwritten by the newest
readings.
Either mode may be selected by pressing:
POWER, SAMPLE and CAL ▲ at the same time, screen will display:
ConF briefly, followed by:
Sch Y or Sch n
Select the required setting by pressing SAMPLE to confirm choice. The display will now show:
SoF Y Stop On Full Yes or SoF n Stop On Full No
Use arrow keys to navigate between the two settings pressing SAMPLE to select the required option.
When the FRAM memory is full the message FULL is shown when the instrument is turned on and if a sample is taken. If Stop
On Full Yes is set, then sampling is prohibited when memory is full.
Various Instrument Modes, Limits and the Optional Alarms may also be set up in a similar way see detailed flow chart that
follows.