EX-5175-EC ENMET Corporation 6 4.0 Operation of the EX-5175-EC It is best to have the EX-5175-EC transmitters powered up and operational for 24 ho
EX-5175-EC ENMET Corporation 7 5.0 Maintenance of the EX-5175-EC CAUTION: Do not open the EX-5175-EC S/T in a classified area. CAUTION: Do Not Att
EX-5175-EC ENMET Corporation 8 Regulator Gas Cylinder Select Menu Output, for Gas Input, for Calibration or Optional Gas Sampling Calibration Cov
EX-5175-EC ENMET Corporation 9 5.2.1 Zero Adjust Zero function is Not applicable for Oxygen sensor/transmitters, go to Section 5.2.2 Gas Span. A Z
EX-5175-EC ENMET Corporation 10 Normal Display Mode 5ppm MENU HOLD the magnet over the MENU switch for 2 – 4 seconds to enter the Maintenan
EX-5175-EC ENMET Corporation 11 5.3 Sensor Replacement CAUTION: Area must be declassified during sensor replacement. Sensors should be replaced wh
EX-5175-EC ENMET Corporation 12 6.0 Replacement Part Numbers ENMET replacement part numbers: Description Part Number For EX-5175-EC p/n Sensor: O
PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instru
Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: 734.761.1270 FAX Number: 734.7
Table of Contents 1.0 INTRODUCTION ...
EX-5175-EC ENMET Corporation 1 1.0 Introduction The ENMET EX-5175-ELECTROCHEMICAL, EC sensor/transmitters (S/T) is a 3-wire, 24 VDC 4-20mA S/T fo
EX-5175-EC ENMET Corporation 2 2.0 Features of the EX-5175-EC See Figure 1 for location of features: Feature Description Display LCD: Indicates th
EX-5175-EC ENMET Corporation 3 3.0 Installation of the EX-5175-EC CAUTION: Area must be declassified during installation. The ENMET EX-5175-EC gas
EX-5175-EC ENMET Corporation 4 3.2 Wiring the EX-5175-EC to a Control Unit CAUTION: Area must be declassified during installation. Run conduit and
EX-5175-EC ENMET Corporation 5 3.3 Start up When the EX-5175-EC transmitter is first powered up, it goes through a series of momentary screens, wh