ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 www.enmet.com Manual Part Number 80002-046 MCN-13-009, 12/09/13

Seite 2 - Table of Contents

PROAIR 2200 ENMET Corporation 7 3.4.1 Relay Contacts Relay contacts are available for each alarm; these are SPDT, rated at 10Amp at 110VAC, and ma

Seite 3 - List of Tables

PROAIR 2200 ENMET Corporation 8 3.4.2 Optional 4-20mA Outputs Isolated 4-20 mA outputs are available for data logging or other purposes. An outpu

Seite 4 - 1.0 Introduction

PROAIR 2200 ENMET Corporation 9 4.0 Operation 4.1 Normal Operation Condition With the PROAIR 2200 installed as described in Section 3, and in clea

Seite 5 - 2.0 Instrument Features

PROAIR 2200 ENMET Corporation 10 4.3 Alarm Latching or Differential Settings An instrument is shipped with the alarms in the latching mode. The a

Seite 6

PROAIR 2200 ENMET Corporation 11 4.6 Operational Menu The operational menu allows the user to:  View alarm set point concentration values  View

Seite 7 - Terminals

PROAIR 2200 ENMET Corporation 12 4.7 Fault Indications 4.7.1 Low Flow Indication A flow sensor is used to furnish a low flow indication. When the

Seite 8 - 3.0 Installation

PROAIR 2200 ENMET Corporation 13 5.0 Maintenance The PROAIR 2200 requires periodic sensor calibration and replacement. Calibration of toxic gas a

Seite 9

PROAIR 2200 ENMET Corporation 14 5.2.2 Maintenance Menu Flow Chart The maintenance menu diagram is shown in Figure 6 Maintenance Menu Flow Chart.

Seite 10 - 3.4.1 Relay Contacts

PROAIR 2200 ENMET Corporation 15 5.3 Calibration for CO, O2, HC and CO2 (Gas Channels) Calibration is the process of setting the instrument up to

Seite 11 - Table 2: Sensor Output

PROAIR 2200 ENMET Corporation 16 Figure 7: Connection of Calibration Gas Cylinder Calibration Process for MOS Sensors The following calibration

Seite 12 - 4.0 Operation

Table of Contents 1.0 INTRODUCTION ...

Seite 13 - 4.5 Display

PROAIR 2200 ENMET Corporation 17 5.3.1B High Cal/SpanGas Adjust A High Cal/Span Gas should only be preformed after a successful Low Cal/ZeroCal ha

Seite 14

PROAIR 2200 ENMET Corporation 18 5.3.3 Set Alarm Points Factory alarm set points are discussed in Section 4.2, See Table 1. To change the alarm p

Seite 15 - 4.7 Fault Indications

PROAIR 2200 ENMET Corporation 19 Example: Set Alarms Flow Chart Displays are examples of Alarms Λ - Indicates alarm triggered on increasing value

Seite 16 - Enter Password

PROAIR 2200 ENMET Corporation 20 5.3.4 Set Alarm Delay The alarms may be set to delay by 1 second increments, up to 255 seconds. Alarm delays are

Seite 17 - Figure 6: P

PROAIR 2200 ENMET Corporation 21 5.3.5 Relay Configuration To change a relay configuration you must enter the maintenance menu. Press the OPTION

Seite 18

PROAIR 2200 ENMET Corporation 22 5.3.6 Failsafe Configuration To change a relay failsafe configuration you must enter the maintenance menu. Pres

Seite 19 - If any

PROAIR 2200 ENMET Corporation 23 5.3.8 Set New Password To change the password, you must enter the maintenance menu. Press the OPTION switch unti

Seite 20

PROAIR 2200 ENMET Corporation 24 5.4 Sensor Replacement 5.4.1 Gas Sensor A Gas sensor must be replaced when it can no longer be calibrated. To re

Seite 21 - 5.3.3 Set Alarm Points

PROAIR 2200 ENMET Corporation 25 After entering a valid password to maintenance menu, see Section 5.2.1, the calibration section is the first menu

Seite 22

PROAIR 2200 ENMET Corporation 26 Figure 7: Connection of Calibration Gas Cylinder Calibration Process for MOS Sensors The following calibratio

Seite 23 - 5.3.4 Set Alarm Delay

List of Figures FIGURE 1: EXTERNAL FEATURES OF THE PROAIR 2200 ...

Seite 24 - 5.3.5 Relay Configuration

PROAIR 2200 ENMET Corporation 27 5.4.4B High Cal/SpanGas Adjust A High Cal/Span Gas should only be preformed after a successful Low Cal/ZeroCal ha

Seite 25 - 5.3.7 Set Output Span Range

PROAIR 2200 ENMET Corporation 28 6.0 Technical Data and Specifications The PROAIR 2200 technical data and specifications: Electrical Power 15 Amp

Seite 26 - 5.3.9 Exit Maintenance Menu

PROAIR 2200 ENMET Corporation 29 7.0 Replacement Part Numbers 7.1 ENMET typical part numbers for sensors: Part number Description 03009-005 De

Seite 27 - 5.4 Sensor Replacement

PROAIR 2200 ENMET Corporation 30 8.0 WARRANTY ENMET warrants new instruments to be free from defects in workmanship and material under normal use

Seite 28 - Default Calibration Points

PROAIR 2200 ENMET Corporation 31 Appendix A: CO Characteristics The Characteristics and Effects of Carbon Monoxide Carbon monoxide is a colorless

Seite 29

PROAIR 2200 ENMET Corporation 32 Appendix B: Gas Ionization Potentials Chemical Name IP (eV) Chemical Name IP (eV) A p-Bromotoluene 8.67 2

Seite 30 - 5.5 Dew Point Sensor

PROAIR 2200 ENMET Corporation 33 Chemical Name IP (eV) Chemical Name IP (eV) (D continued) F Decaborane 9.88 2-Furaldehyde 9.21 Diazom

Seite 31

PROAIR 2200 ENMET Corporation 34 Chemical Name IP (eV) Chemical Name IP (eV) M (P continued) 2-Methyl furan 8.39 2-Pentanone 9.38 2-Me

Seite 32 - 7.0 Replacement Part Numbers

PROAIR 2200 ENMET Corporation 35 Chemical Name IP (eV) Chemical Name IP (eV) (T continued) W Trimethyl amine 7.82 Water 12.59 Tripropy

Seite 33 - 8.0 WARRANTY

PO Box 979 680 Fairfield Court Ann Arbor, Michigan 48106-0979 734.761.1270 Fax 734.761.3220 Returning an Instrument for Repair ENMET instrumen

Seite 34

PROAIR 2200 ENMET Corporation 1 1.0 Introduction The PROAIR 2200 is a compressed air monitoring instrument that measures and detects certain haza

Seite 35 - ENMET Corporation

Repair Return Form Mailing Address: ENMET Corporation PO Box 979 Ann Arbor, Michigan 48106 Phone Number: 734.761.1270 FAX Number: 734.

Seite 36

PROAIR 2200 ENMET Corporation 2 2.0 Instrument Features 2.1 Exterior Features The exterior of the instrument is shown in Figure 1. The exterior

Seite 37

PROAIR 2200 ENMET Corporation 3 Figure 1: External Features of the PROAIR 2200 * NOTE: Typical gas reading & alarm locations, depending on

Seite 38

PROAIR 2200 ENMET Corporation 4 2.3 Circuit Board Features The Display Panel is hinged on the left and is released by unscrewing the 2 screws loc

Seite 39

PROAIR 2200 ENMET Corporation 5 3.0 Installation 3.1 Mounting of Instrument The PROAIR 2200 should be located near the pipe or tank containing the

Seite 40 - Repair Return Form

PROAIR 2200 ENMET Corporation 6 3.2 Sample Air Supply Tap the pipe or tank containing the breathing air and use appropriate fittings to connect th